gzspark61 medical school senior
Joined: 11 May 2008 Posts: 33 Location: northren virginia |
Posted: Tue Jun 23, 2009 1:46 pm Post subject: 2200 scanner glass or platen becomes dirty (214) |
|
|
Screen shows please wait on start up then goes directly to error.
Resat all connectors inside and out.
Made sure all boards are connected tightly.
Glass and platen clean.
I'm at a point I put a call into KIP and waiting at customers.
Anyone run into this before?
Thanks |
|
|
|
|
|
|
|
|
|
|
CHILLIN doctor
Joined: 05 Oct 2006 Posts: 950
|
Posted: Tue Jun 23, 2009 4:30 pm Post subject: |
|
|
KIP has Tech TIP on this that might help. TIP-K98-002. Let me know if you need it and I will e-mail it to you. |
|
|
|
|
|
|
|
|
|
|
Brainz resident
Joined: 28 Oct 2008 Posts: 210 Location: World |
Posted: Wed Jun 24, 2009 1:25 am Post subject: |
|
|
This Kip tip has personally never worked for me. Everytime I have this issue, it ends up being the CPU board, and this board really affects your pocket |
|
|
|
|
|
|
|
|
|
|
gzspark61 medical school senior
Joined: 11 May 2008 Posts: 33 Location: northren virginia |
Posted: Wed Jun 24, 2009 3:47 pm Post subject: |
|
|
Thank for the offer CHILLIN but Brainz is right it is the CPU....OUCH!!! |
|
|
|
|
|
|
|
|
|
|
cezwhat resident
Joined: 24 Aug 2005 Posts: 221 Location: On the Prairie |
Posted: Thu Jun 25, 2009 11:31 am Post subject: |
|
|
I had this same issue 2 weeks ago I was able to force start up ( hold the start button and press 1) on power up and update the firmware. It saved me a ton time and money. _________________ Giv'em hell before they give it to you! |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|